Reply To: SM200B receive problem in low frequency band

Forums › SM Series Discussions › SM200B receive problem in low frequency band › Reply To: SM200B receive problem in low frequency band

#248179
AndrewAndrew
Moderator

mmark,

There are a couple scaling factors you might need to consider when FFT’ing the I/Q data from our API.

– If you are requesting 16-bit I/Q data from our API, and you want to measure dBm, then you will need to get the I/Q correction and apply it via the equations in our API manual. If you are requested 32-bit floating point data (recommended), then the data is already scaled.

– The FFT itself. I would recommend scaling the Fwd FFT by N. What additional scaling you need to apply is dependent on which FFT library you are using. Some use scaling by default, some dont.

– If you are using a window function, you want to normalize the window before windowing the data.

– Once you have the above resolved, the output of the FFT can be converted to dBm using the equation

10*log10(I^2 + Q^2)